Lens Meter Maintenance and Trouble-shooting
Home News
Lens Meter Maintenance and Trouble-shooting

Lens Meter Maintenance and Trouble-shooting

October 22, 2024
Maintenance
(1)Don’t put machine under sunshine.
(2)Don’t put heavy part on machine, which will damage to machine.
(3)To keep clean both machine and air around.
(4)To keep machine clean. Don’t use the organic solution like volatile liquid, diluent or benzol to machine.
(5)No operation for long time, to cut off power supply and pull out the plug.
(6)Please prepare a dust-proof cover and cover the instrument when it is free in order to reduce the influence of dust.
(7)Please set screen protection when the instrument is in use, if it is free of test the screen can shut automatically.
(8)Clean the protecting lens regularly. Take down the bracket first and clear the dust or oil with soft brush or lens paper.
Warning:
鈼?nbsp;Don’t scratch the protecting lens, otherwise the precision will be reduced.
鈼?nbsp;lens bracket must place well to avoid measurement mistake.

 

Trouble shooting

NoFaultReasonSolution
1No reaction from screen when turn on
machine
There power wires are not connected well, or turn off machine not correctlyTo check the wires if in good condition, and cut off the power then turn on it after a while.
2The screen doesn’t show anything suddenly.The screen protection has been set.To touch screen and enter into Menu to set a reasonable time for screen protection.
3When power is on, but also can’t switch on the monitor.monitor auto off function is on or monitor brightness volume is lowpress any switch to resume the screen or adjust the brightness volume
4The data is not 0 when there is no lens.The light is too strong or the protecting glass is dirtyTo avoid the strong light or clean the glass
5Not reaction when press the printNo measured data saved.The data should be saved then can be printed.
6The marks on lens not clearThere is problem with the dot pen.To change the dot pen
7Measured data preciselessNeed to calibrate the data againEnter into the interface for calibration